172-65639M-01 (EF200) 13 Apr 2018
19
4.2 Wiring Diagrams
EF200 Only, Connecting Power Supply
Frequency (Pulse) Output for an Electronic Counter or PLC
Caution!
The residual voltage of pulse output is 2 V at load current of 2 mA or less,
or 8 V at 10 mA. Make sure to select an electronic counter or PLC
conforming to these specifications or an instrument that can be set to “Low”
level.
Current Output for an Analog Data Receiver
NOTE: Shielded cable recommended for current output.
Power
Supply
+24V
+
–
Electric Counter
or PLC
(Open collector input)
Power
Supply
+24V
+
–
Power
Supply
+24V
+
–
+
–
Data Receiver
(Analog 4 – 20 mA
current input)
Figure 15
Power Supply DC24V
Figure 17
Analog current output
connection
Figure 16
Pulse output to electronic
counter or PLC
Caution!