XD 7500
Operation
69
ba77191e01 - 71310060
04/2018
The related formula f(x) and cor-
relation coefficient R
2
are dis-
played above the diagram.
12
If required, select a different line
type for the line adjustment with
[Curve type]
.
-
Polygon line
-
Straight line
-
Parabola
13
If required, enter different mea-
sured value limits with
[Meas.
range]
.
-
Lower limit
-
Upper limit
14
Using
[Next]
, complete the editing
of the calibration line and proceed
to the next page.
The timers and MCheck data
linked to the method are dis-
played.
15
If necessary, enter intervals for up
to 4 timers.
16
If necessary, enter method check
parameter
MCheck target value
and
MCheck tolerance
.
17
If necessary, set the number from
how many single measurements
the documented measured value
will be calculated.
18
If necessary, specify whether a
reagent blank value is required.
19
If necessary, set whether a user
calibration is possible and/or
required.
20
Complete the programming of the
method with
[Complete]
.
The method is programmed and
selected for measuring.
Edit method
16.01.17 9:52
f(x)=0.73x
R
2
= 1.000
Meas. range: 0.050 - 1.000 mg/l
Concentration [mg/l NO2-N]
Back
Curve type
Meas. range
Next
Absorban
ce
Edit method
16.01.17 9:52
Timer 0
00:00:00
Timer 1
00:00:00
Timer 2
00:00:00
Timer 3
00:00:00
MCheck target value
1.00 mg/l
MCheck tolerance
0.10 mg/l
Required measurements
1
Reagent blank value required
No
User calibration possible
No
User calibration required
No
Back
Complete