Operation
XD 7500
52
ba77191e01 - 71310060
04/2018
The photometer is ready for mea-
surement.
1
Use
<ZERO·BLANK>
to open
the
Adjust
selection list.
or
Open the settings menu with
[Setup]
.
2
Select and confirm
Reagent blank
.
The window for the measurement
of the reagent blank value pops
up.
The data of the last measurement
appears in the measured value
display.
3
Insert the cell with the blank sam-
ple.
The first single measurement for
the reagent blank value takes
place.
The following data is displayed as
the result:
– The measured absorbance
from the (last) single measure-
ment.
– The median from all single mea-
surements carried out up to
now.
Concentration
16.01.17 9:52
To start measurement,
insert cell or press <START/ENTER>
NH
4
-N
16 mm
0.02 - 2.50 mg/l
Setup
Method list
Citation form
Unit
Concentration
16.01.17 9:52
NH
4
-N
16 mm
0.02 - 2.50 mg/l
Setup
Method list
Citation form
Unit
Adjust
Zero adjustment
Reagent blank value
Calibrate the method
Reagent blank value
16.01.17 9:52
To start measurement,
insert cell or press <START/ENTER>
NH
4
-N
16 mm
0.02 - 2.50 mg/l