XD 7500
Operation
59
ba77191e01 - 71310060
04/2018
Measuring with user
calibration
Calibrate the method
16.01.17 9:52
User calibration:
Protocol ID:
2
Date:
16.01.2012
User:
Admin
Curve type:
Straight line
Correction:
105%
End
Calibration
Delete
New
Selecting a method manually
(see Section 4.5.5).
If data for zero adjustment,
reagent blank value or a user cali-
bration are already existing, the
photometer informs you of this.
You can accept or discard the
existing values.
If the existing user calibration
should not be utilized, a query with
further options pops up:
-
Use default calibration
The existing user calibration is
erased. The following measure-
ments are done with the original
calibration that was stored with
the method.
-
Carry out user calibration
The existing user calibration is
erased. The process for a new
user calibration is started.
-
Cancel
The existing user calibration is
retained. The previous query is
displayed.
<HOME>
Concentration
Concentration
16.01.17 9:52
[Cal][BV/2c][ZERO 10.01.2012 11:08]
To start measurement,
insert cell or press <START/ENTER>
Setup
Method list
Citation form
Unit
User calibration
A calibration dated xxx is available for this
method. Should it be used?
Yes
No