Operation
XD 7500
136
ba77191e01 - 71310060
04/2018
During the check the test is carried out with the standard solution as the
sample while the other conditions are the same. The photometer com-
pares the result with the nominal value while taking the tolerance into
account.
The steps are described in detail below.
General MCheck
settings
Switching on
MCheck monitoring
for a method
1
Select and confirm
Mode
.
The
Mode
selection field pops up.
2
Select and confirm
Weeks
or
Mea-
surements
.
For all methods, the MCheckinter-
vals are entered either in weeks or
number of measurements.
3
Accept the general settings with
[Apply]
.
When the mode (
Weeks
or
Measurements
) is changed, all
MCheck intervals are reset to the preset values.
<HOME>
DeviceCheck
–
MCheck setup
MCheck setup
16.01.17 9:52
Mode
Weeks
Method ...
Method list
<HOME>
DeviceCheck
–
MCheck setup
–
Method
...