XD 7500
Operation
25
ba77191e01 - 71310060
04/2018
4
Operation
4.1
Switching on or off the photometer
Switching on
Starting the
Self test
Self test
During the self-test, all cells must be removed and the cell shaft cover closed.
The self-test runs in the background and may take some minutes.
1
Switch the photometer on with
<ON/OFF>
.
The display shows
– the
Self test
dialog (if the user
management is not active).
or
– the
Login
dialog
(with activated user manage-
ment).
With activated user management:
2
Login
Enter user name and password or
register as a guest (see Section
4.16.4).
Then the photometer displays the
Self test
dialog.
Self test
16.01.17 9:52
Please make sure no cell is inserted and the cover is closed.
Then press <START/ENTER>
Setup
Info
Login
16.01.17 9:52
Enter user name
Administrator
3
Remove all cells and close the cell
shaft cover.
4
Start the self-test with
<START·ENTER>
.
The photometer carries out the
self-test.
Self test
16.01.17 9:52
Please make sure no cell is inserted and the cover is closed.
Then press <START/ENTER>