3900 Serial I/O
Page 35
Read Generated PPMw
?PW
Description
Returns the current value of the PPMw, Parts Per Million by Weight,
being generated.
Prerequisites
Firmware Ver 2.0 or newer
Syntax
?PW
Parameters
None
Remarks
Data Returned
The value returned is the PPMw value calculated from the current
generator values for Ts, Ps, Tt, Pt and the molecular weight of the gas
type being used.
Examples
The
?PW
command is used to request the current PPMw value:
computer sends
?PW
<CR>
3900 responds
1546.249
<CR><LF>
The generator is currently generating 1546.249 parts per million by
weight.
See Also
?