3900 Serial I/O
Page 37
Read Generated %RH
?RH
Description
Returns the current value of the %RH being generated.
Prerequisites
Firmware Ver 2.0 or newer
Syntax
?RH
Parameters
None
Remarks
Data Returned
The value returned is the %RH value calculated from the current
generator values for Ts, Ps, Tt, and Pt as defined by the RH calculation
method.
Examples
The
?RH
command is used to request the current %RH value:
computer sends
?RH
<CR>
3900 responds
10
<CR><LF>
The generator is currently generating 10% RH.
See Also
?, ?WM, WM=