3900 Serial I/O
Page 38
RH=
Change RH Setpoint
Description
Changes the RH setpoint to a given value.
Prerequisites
None
Syntax
RH=
setpoint
Parameters
setpoint
The
setpoint
value is the %RH that the system should generate.
Remarks
Control Mode Changed
Sending
RH=
setpoint
also changes the control mode to %RH (mode 4).
In this mode Frost Point, Dew Point, PPMv, PPMw and Saturation
Pressure setpoints vary as system temperatures and pressures change.
Data Returned
None. The only response is a carriage return/linefeed terminator.
Examples
The
RH=
setpoint
command is used to set the %RH setpoint to 10.
computer sends
RH=10
<CR>
3900 responds
<CR><LF>
The
<CR><LF>
response signifies completion of the command.
See Also
?SP