3-21
An incorrect code prevents access and returns to step 3. A correct code results in the
display of calibration coefficients.
4) Using the [NEXT] and [PREV] keys, view any or all of the remaining coefficient and
parameter screens.
5) To edit a particular displayed value, press [EDIT]. The cursor begins flashing at the left
of the first parameter displayed.
Using [-/+], ['E'], arrow keys, and the numeric keys as necessary, change any or all
displayed values as desired. Then press <ENTER>.
6) After Editing or Viewing, press [DONE]. Then from the next menu, press [DONE]
again. The system reinitializes back to the Control/Display screen.
SATURATION TEMP, Ts
Zero
[-.877401
]
Span
[ 9.78206E-03]
Lin
[-6.07702E-09]
Avg
[ 10
]
Last Cal Date 03/17/10
EDIT
NEXT
PREV
DONE
MASS FOLW RATE
Zero
[ 0
]
Span
[ 4E-04
]
Lin
[ 0
]
Avg
[ 10
]
Last Cal Date 03/14/10
EDIT
NEXT
PREV
DONE
MASS FOLW RATE
Zero
[ _0
]
Span
[ 4E-04
]
Lin
[ 0
]
Avg
[ 10
]
Last Cal Date 03/14/10
-/
/+
'E'