4-14
Note -
For all transducers, ambient pressure may be used for the low
pressure calibration point.
Note -
If a mistake is made during reference pressure entry, pressing
the [OOPS] key cancels the data entry mode, leaving all values
for that point unchanged.
13) Apply the mid range pressure and watch the displayed value. Once stable, press the
[MID PRES] key and enter the reference pressure.
14) Apply the upper recommended pressure and watch the displayed value. Once stable,
press the [HIGH PRES] key and enter the reference pressure.
15) Press the [CALC COEF] key. All coefficients for the marked transducer will be
calculated, and appear on the LCD display.
Count psi A
*Low Range
15538
30.11
Hi Range
1513
12.19
Ext Test 7385
12.19
Supply
29806
250.9
Low
Mid
High
12.212
29.989_
50
-/
/+
OOPS
Count psi A
*Low Range
15538
30.11
Hi Range
1513
12.19
Ext Test 7385
12.19
Supply
29806
250.9
Low
Mid
High
12.212
29.989_
50
-/
/+
OOPS
LOW RANGE PRESSURE, Ps
Zero [-.609974 ]
Span [ 1.96204E-03]
Lin [ 5.12663E-09]
PRNT
COEF
EXIT
QUIT
SAVE
QUIT