3900 Serial I/O
Page 16
FP=
Change Frost Point Setpoint
Description
Changes the Frost Point setpoint to a given value.
Prerequisites
None
Syntax
FP=
setpoint
Parameters
setpoint
The
setpoint
value is the Frost Point, in °C, that the system should
generate.
Remarks
Control Mode Changed
Sending
FP=
setpoint
also changes the control mode to Frost Point
(mode 0). In this mode PPMv, PPMw, %RH and Saturation Pressure
setpoints vary as system temperatures and pressures change.
Data Returned
None. The only response is a carriage return/linefeed terminator.
Examples
The
FP=
setpoint
command is used to set the Frost Point setpoint to
-10 °C.
computer sends
FP=-10
<CR>
3900 responds
<CR><LF>
The
<CR><LF>
response signifies completion of the command.
See Also
?SP