![Thunder Scientific series 3900 Скачать руководство пользователя страница 80](http://html1.mh-extra.com/html/thunder-scientific/series-3900/series-3900_operation-and-maintenance-manual_1113431080.webp)
4-21
12) Press [CALC COEF] to calculate the new flow meter coefficients. These new
coefficients appear on the display.
13) If a printer is attached, a calibration record of the flow points and calculated
coefficients is printed. Press the [PRNT COEF] key if additional printouts are
desired. This is the only opportunity to print this report. Upon leaving this screen,
the data used for report generation is lost (with the exception of the calculated
coefficients being displayed.)
14) To save the coefficients to non-volatile memory, press [SAVE QUIT]. To discard
these new coefficients and revert to the previous ones, press [EXIT QUIT].
15) At the calibration menu, press [DONE]. At the next menu, press [DONE] again. The
system reinitializes to the Control / Display screen.
16) To print a calibration report, refer to section 4.2.5.
MASS FLOW RATE
Zero [-1.56852E-03]
Span [ 3.92120E-04]
Lin [ 2.62802E-09]
PRNT
COEF
EXIT
QUIT
SAVE
QUIT