![Thermo 42i Скачать руководство пользователя страница 69](http://html.mh-extra.com/html/thermo/42i/42i_instruction-manual_1104217069.webp)
Operation
Instrument Controls Menu
Thermo Electron Corporation
Model 42
i
Low Source Instruction Manual
3-41
•
In the Main Menu, choose Instrument Controls > Datalogging Settings
> Configure Datalogging >
Memory Allocation %
.
•
Use
and
to scroll through a list of choices.
•
Press
to set the percentage for both record types and proceed to
the erasure warning screen.
•
Press
to confirm change.
Data Treatment
The Data Treatment screen is used to select the data type for the selected
record: whether the data should be averaged over the interval, the minimum
or maximum used, or the current value logged. Data treatment doesn’t
apply to all data, just to the concentration measurement. All other data
points log the current value at the end of the interval.
•
In the Main Menu, choose Instrument Controls > Datalogging Settings
> Configure Datalogging >
Data Treatment
.
•
Use
and
to scroll through a list of choices.
•
Press
to set the data type.
SET PERCENT SRECS:
SET PERCENT SRECS:
CURRENTLY: 50 %
CURRENTLY: 50 %
SET TO: 60 % ?
SET TO: 60 % ?
THIS WILL ERASE ALL LOGS!
x
PRESS
¨
TO CONFIRM CHANGE
ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
RANGE AVG DIAGS ALARM
SET LREC DATA TYPE:
CURRENTLY: AVG
SET TO: CUR ?
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
Содержание 42i
Страница 4: ...WEEE Compliance WEEE Compliance Thermo Electron Corporation ...
Страница 14: ...Tables xiv Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 28: ...Installation Startup 2 6 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 128: ...Operation Password 3 100 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 146: ...Calibration Zero and Span Check 4 18 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 154: ...Preventive Maintenance Pump Rebuilding 5 8 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 224: ...Servicing Service Locations 7 48 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 238: ...System Description I O Components 8 14 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 246: ...Optional Equipment Mounting Options 9 8 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...