System Description
I/O Components
Thermo Electron Corporation
Model 42
i
Low Source Instruction Manual
8-11
Digital Inputs
Sixteen digital inputs are available which may be programmed to signal
instrument modes and special conditions including:
•
NO Measure Mode
•
NO
x
Measure Mode
•
Zero Gas Mode
•
Span Gas Mode
The actual use of these inputs will vary based on analyzer configuration.
The digital inputs are TTL level compatible and are pulled up within the
analyzer. The active state can be user defined in firmware.
Serial Ports
Two serial ports allow daisy chaining so that multiple analyzers may be
linked using one PC serial port.
The standard bi-directional serial interface can be configured for either
RS-232 or RS-485. The serial baud rate is user selectable in firmware for
standard speeds from 1200 to 19,200 baud. The user can also set the data
bits, parity, and stop bits. The following protocols are supported:
•
C-Link
•
Streaming Data
•
Modbus Slave
The Streaming Data protocol transmits user-selected measurement data via
the serial port in real-time for capture by a serial printer, data logger, or PC.
RS-232 Connection
A null modem (crossed) cable is required when connecting the analyzer to
an IBM Compatible PC. However, a straight cable (one to one) may be
required when connecting the analyzer to other remote devices. As a general
rule, when the connector of the host remote device is female, a straight cable
is required and when the connector is male, a null modem cable is required.
Data Format:
1200, 2400, 4800, 9600, 19200, 38400, 57600, or 115200 BAUD
8 data bits
Содержание 42i
Страница 4: ...WEEE Compliance WEEE Compliance Thermo Electron Corporation ...
Страница 14: ...Tables xiv Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 28: ...Installation Startup 2 6 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 128: ...Operation Password 3 100 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 146: ...Calibration Zero and Span Check 4 18 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 154: ...Preventive Maintenance Pump Rebuilding 5 8 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 224: ...Servicing Service Locations 7 48 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 238: ...System Description I O Components 8 14 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 246: ...Optional Equipment Mounting Options 9 8 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 314: ......