![Thermo 42i Скачать руководство пользователя страница 269](http://html.mh-extra.com/html/thermo/42i/42i_instruction-manual_1104217269.webp)
C-Link Protocol Commands
Datalogging
Thermo Electron Corporation
Model 42
i
Low Source Instruction Manual
B-21
These commands set the long and short records logging period to
value
in
minutes. The example below sets the long record logging period to 15
minutes.
Send:
set lrec per 15
Receive:
set lrec per 15 ok
no of lrec
no of srec
These commands report the number of long and short records stored in the
long and short records memory. The example below shows that 50 long
records have been stored in the memory.
Send:
no of lrec
Receive:
no of lrec 50 recs
malloc lrec
malloc srec
These commands report the currently set memory allocation for long and
short records in percent of total memory.
Send:
malloc lrec
Receive:
malloc lrec 10%
set malloc lrec
value
set malloc srec
value
value
= 0 to 100
These commands set the percent of memory space allocated for long and
short records to
value
, where
value
is a floating-point number representing
percent. The example below sets the memory allocation for long records to
10.
Note
Issuing these commands will clear all the logging data memory. All the
existing records should be retrieved using appropriate commands, if
required.
Send:
set malloc lrec 10
Receive:
set malloc lrec 10 ok
set copy sp to lrec
set copy sp to srec
set copy sp to stream
These commands copy the current selections in scratch pad (sp) into the
long record, short record, or streaming data list. The example below copies
the current list in scratch pad into the long records list.
Содержание 42i
Страница 4: ...WEEE Compliance WEEE Compliance Thermo Electron Corporation ...
Страница 14: ...Tables xiv Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 28: ...Installation Startup 2 6 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 128: ...Operation Password 3 100 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 146: ...Calibration Zero and Span Check 4 18 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 154: ...Preventive Maintenance Pump Rebuilding 5 8 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 224: ...Servicing Service Locations 7 48 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 238: ...System Description I O Components 8 14 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 246: ...Optional Equipment Mounting Options 9 8 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 314: ......