Troubleshooting
Troubleshooting Guides
Thermo Electron Corporation
Model 42
i
Low Source Instruction Manual
6-3
Unstable NO or NO
2
source
Replace.
Clogged capillaries
Unplug power cord. Clean or
replace capillary.
Clogged sample air filter
Replace filter element.
Excessive noise
Defective or low sensitivity PMT
Unplug power cord. Remove
PMT. Install known good
PMT. Plug in power cord.
Check performance.
Defective input board
Replace board.
Defective cooler
Check temperature (less
than
-2 °C at T
amb
= 25 °C).
Non-linear response
Incorrect calibration source
Verify accuracy of multipoint
calibration source gas.
Leak in sample probe line
Check for variable dilution.
Excessive response
time
Partially blocked sample capillary
Unplug power cord. Clean or
replace capillary.
Hang up/blockage in sample filter
Change element.
Improper converter
operation
Questionable calibration gas
Verify accuracy.
Converter temperature too high or
too low
Temperature should be
approximately 325 °C.
Low line voltage
Check to see if line voltage
is within specifications.
Molybdenum consumed
Replace Molybdenum
converter cartridge.
Table 6-2.
Troubleshooting - Alarm Messages
Alarm Message
Possible Cause
Action
Alarm - Cooler Temp
High
Check fan operation
Replace defective fan.
Check fan filter
Clean or replace foam filter.
Cooler reads 80 °C
Bad unit
Replace cooler.
Cooler does not hold set point of -3
°C
Replace cooler –
thermoelectric module
inside cooler failed.
Table 6-1.
Troubleshooting - General Guide, continued
Malfunction
Possible Cause
Action
Содержание 42i
Страница 4: ...WEEE Compliance WEEE Compliance Thermo Electron Corporation ...
Страница 14: ...Tables xiv Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 28: ...Installation Startup 2 6 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 128: ...Operation Password 3 100 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 146: ...Calibration Zero and Span Check 4 18 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 154: ...Preventive Maintenance Pump Rebuilding 5 8 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 224: ...Servicing Service Locations 7 48 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 238: ...System Description I O Components 8 14 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 246: ...Optional Equipment Mounting Options 9 8 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 314: ......