C-Link Protocol Commands
Datalogging
B-18
Model 42
i
Low Source Instruction Manual
Thermo Electron Corporation
erec
This command returns a brief description of the main operating conditions
at the time the command is issued (i.e. dynamic data). The example below
shows a typical response. The format is defined by the current settings of
“format” and “erec format” commands. For details on how to decode the
flag fields within these records, see Figure B-1 in the “flags” command.
Send:
erec
Receive:
erec
10:11 04-06-05 flags DD008000 no 0.000 nox 0.000 no2 0.000 1 lono
147.500 lonox 0.000 lono2 0.000 1 pmtv 805.491 tempal 1 pres 172.278
pcal 150.000 smplf 0.000 ozonf 0.050 hiavgtime 10 loavgtime 10 nobkg
0.000 noxbkg 0.000 nocoef 1.000 noxcoef 1.000 no2coef 1.000 lonocoef
1.000 lonoxcoef 1.000 lono2coef 1.000 norange 100000.000 noxrange
100000.000 no2range 100000.000 lonorange 100000.000 lonoxrange
100000.000 lono2range 100000.000
lrec
srec
lrec
xxxx yy
srec
xxxx yy
lrec
aa:bb oo-pp-qq yy
srec
aa:bb oo-pp-qq yy
xxxx
= the number of past records
yy
= the number of records to return (1 to 10)
aa
= hours (01 to 24)
bb
= minutes (01 to 59)
oo
= month (01 to 12)
pp
= day (01 to 31)
= year
These commands output long or short records and dynamic data. The
output format is determined by the “set lrec format
”,
and “set srec format
”
commands. The logging time is determined by the “set lrec per” and “set
srec per” commands. In dual range, the long records and short records
contain the high and low NO and NO
x
concentrations. In single range the
low NO and low NO
x
values are set to 0 and the high NO and high NO
x
are used. In NO or NO
x
only mode, the pertinent high value used, other
concentrations are set to 0. Concentrations are stored in either ppb or
µ
g/m
3
.
In the following example, there are 740 long records currently stored in
memory. When the command lrec 100 5 is sent, the instrument counts back
100 records from the last record collected (record 740), and then returns 5
records: 640, 641, 642, 643, and 644. For details on how to decode the flag
fields within these records, see Figure B-1 in the “flags” command.
Содержание 42i
Страница 4: ...WEEE Compliance WEEE Compliance Thermo Electron Corporation ...
Страница 14: ...Tables xiv Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 28: ...Installation Startup 2 6 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 128: ...Operation Password 3 100 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 146: ...Calibration Zero and Span Check 4 18 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 154: ...Preventive Maintenance Pump Rebuilding 5 8 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 224: ...Servicing Service Locations 7 48 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 238: ...System Description I O Components 8 14 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 246: ...Optional Equipment Mounting Options 9 8 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 314: ......