Servicing
Analog Output Testing
Thermo Electron Corporation
Model 42
i
Low Source Instruction Manual
7-31
Figure 7-17.
Rear Panel Analog Voltage Output Pins
2. From the Main Menu,
press
to scroll to Diagnostics, > press
>
to scroll to Test Analog Outputs, and press
.
The Test Analog Outputs screen appears.
3. Press
to scroll to the channel (Voltage Channel 1-6)
corresponding to the rear panel terminal pin where the meter is
connected, and press
.
Table 7-3.
Analog Output Channels and Rear Panel Pin Connections
Channel
Pin
1
14
2
33
3
15
4
34
5
17
6
36
Ground
16, 18, 19, 35, 37
Содержание 42i
Страница 4: ...WEEE Compliance WEEE Compliance Thermo Electron Corporation ...
Страница 14: ...Tables xiv Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 28: ...Installation Startup 2 6 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 128: ...Operation Password 3 100 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 146: ...Calibration Zero and Span Check 4 18 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 154: ...Preventive Maintenance Pump Rebuilding 5 8 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 224: ...Servicing Service Locations 7 48 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 238: ...System Description I O Components 8 14 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 246: ...Optional Equipment Mounting Options 9 8 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 314: ......