Calibration
Calibration
4-6
Model 42
i
Low Source Instruction Manual
Thermo Electron Corporation
6. Compute t
R
as:
7. Verify that t
R
< 2 minutes. If not, select a reaction chamber with a
smaller V
RC
.
8. Compute the diluent air flow rate as:
9. If F
O
turns out to be impractical for the desired system, select a reaction
chamber having a different V
RC
and re-compute F
D
and F
O
.
Adjusting Instrument Gain
Use the following procedure to adjust the instrument gain.
Note
The NO channel should be calibrated first and then calibrate the NO
x
channel.
For detailed information about the menu parameters and the icons used in
these procedures, see the “Operation” chapter.
1. Determine the GPT flow conditions required to meet the dynamic
parameter specifications as indicated in “Dynamic Parameter
Specifications for Gas Phase Titrator” earlier in this chapter.
2. Adjust the GPT diluent air and O
3
generator air flows to obtain the
flows determined in “Dynamic Parameter Specifications for Gas Phase
Titrator” earlier in this chapter. The total GPT airflow must exceed the
total demand of the analyzer. The Model 42
i
Low Source requires
approximately 700 cc/min of sample flow, and a total GPT airflow of at
least 1.5 liters/min is recommended.
a. Allow the analyzer to sample zero air until the NO, NO
x
and NO
2
responses stabilize.
b. After the responses have stabilized, from the Main Menu, choose
Calibration >
Calibrate NO Background
.
c. Press
to set the NO reading to zero.
F
+
F
V
=
t
NO
O
RC
R
F
-
F
-
F
=
F
NO
O
T
D
Содержание 42i
Страница 4: ...WEEE Compliance WEEE Compliance Thermo Electron Corporation ...
Страница 14: ...Tables xiv Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 28: ...Installation Startup 2 6 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 128: ...Operation Password 3 100 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 146: ...Calibration Zero and Span Check 4 18 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 154: ...Preventive Maintenance Pump Rebuilding 5 8 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 224: ...Servicing Service Locations 7 48 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 238: ...System Description I O Components 8 14 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 246: ...Optional Equipment Mounting Options 9 8 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 314: ......