![Thermo 42i Скачать руководство пользователя страница 191](http://html.mh-extra.com/html/thermo/42i/42i_instruction-manual_1104217191.webp)
Servicing
Photomultiplier Tube Replacement
Thermo Electron Corporation
Model 42
i
Low Source Instruction Manual
7-15
7. Loosen four captive screws holding cooler to floor plate and remove the
cooler assembly with the reaction chamber.
Note
If only the cooler is being replaced, remove the PMT and reaction
chamber from the old cooler and install them on the new cooler.
Fasten knurled fittings on reaction chamber finger tight.
Make sure that the heat shrink covered tubing between the reaction
chamber and the converter is light tight at the connections.
8. Install new cooler by following previous steps in reverse.
9. Re-install the measurement bench. Refer to “Removing the
Measurement Bench and Lowering the Partition Panel” in this chapter.
Photomultiplier Tube
Replacement
Use the following procedure to replace the PMT tube.
Equipment Required:
Photomultiplier tube and PMT base
Nut driver, 5/16-inch
Flat blade screwdriver
Philips screwdriver, small
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Disconnect the high voltage cable from the PMT power supply and
unplug the BNC cable from the Input Board.
3. Remove six external screws holding PMT cover plate and the four
screws holding the PMT shroud to the panel and remove the PMT
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
Содержание 42i
Страница 4: ...WEEE Compliance WEEE Compliance Thermo Electron Corporation ...
Страница 14: ...Tables xiv Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 28: ...Installation Startup 2 6 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 128: ...Operation Password 3 100 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 146: ...Calibration Zero and Span Check 4 18 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 154: ...Preventive Maintenance Pump Rebuilding 5 8 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 224: ...Servicing Service Locations 7 48 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 238: ...System Description I O Components 8 14 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 246: ...Optional Equipment Mounting Options 9 8 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 314: ......