Preventive Maintenance
Pump Rebuilding
Thermo Electron Corporation
Model 42
i
Low Source Instruction Manual
5-5
3. Re-install the filters and fan guards.
Pump Rebuilding
Use the following procedure to rebuild the pump (
). To replace
the pump. See “Pump Replacement” in the “Servicing” chapter.
Equipment Required:
Pump Repair Kit
Allen Wrench, 3 mm
Wrench, 9/16-inch
Needlenose Pliers
1. Turn instrument off, unplug the power cord, and remove the cover.
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component. For more information
about appropriate safety precautions, see the “Servicing” chapter.
Содержание 42i
Страница 4: ...WEEE Compliance WEEE Compliance Thermo Electron Corporation ...
Страница 14: ...Tables xiv Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 28: ...Installation Startup 2 6 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 128: ...Operation Password 3 100 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 146: ...Calibration Zero and Span Check 4 18 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 154: ...Preventive Maintenance Pump Rebuilding 5 8 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 224: ...Servicing Service Locations 7 48 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 238: ...System Description I O Components 8 14 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 246: ...Optional Equipment Mounting Options 9 8 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 314: ......