10 Initial commissioning
03/2007
45
DA0621_a/
Figure 14: Footprint and dimensions, VARIOKLAV 25T, seen from above and from
the side
♦
Choose a stable and level area for deploying the sterilizer (e.g. a table or
workbench). Make sure that this area is exactly level in all directions.
♦
When installing the sterilizer, keep a distance of at least 100 mm from the
wall and from other devices on all sides. The unit must be freely
accessible from above (open to the top).
♦
Place the insertion tray and, if necessary, the rack system inside the
sterilizer.
10.4.1 Connecting the steam sterilizer to the
disposal system
:
CAUTION
During the heating phase, contaminated exhaust air or waste
water may be emitted from the sterilizer to the on-site disposal
system.
Remedy: Expansion option Type FA exhaust air filtration.
Exhaust air, spent steam, condensate (approximately 0.5 to 1 litres per batch) and
waste water are generated during the whole sterilization process.
The outlets situated at the back of the sterilizer must be connected with a disposal
system.
Disposal system: Minimum requirements
The disposal line must be designed for taking in steam or for high temperatures.
Hot water (approximately 100 °C) can flow a short way through the waste-water
channel. (The plastic pipes and channels found on-site are usually not suitable for
this temperature range.)
580 mm
Öffnungs-
winkel 95°
380 mm
1000 mm
680 mm
620 mm
53
0 m
m