6 Programs
03/2007
31
DA0621_a/
6.6 Setting the timer
If you do not want to start sterilization right away, you may set the timer.
When the timer is set, the acoustic signal is automatically deactivated. You can
thus perform sterilization overnight or over the weekend. When you start work the
next (working) day, your sterilized items are ready and waiting for you.
Ö
Setting the starting time and turning on the timer
To set the starting time on the timer, you must observe the current values for
date and time. Chapter 9 describes how you can set the date and time.
Select the desired starting time, and the date if required:
♦
Press and hold the Timer key (23).
♦
Press the START/STOP key (19).
♦
Release both keys (23, 19).
The cycle display (16) now shows the timer date or the timer time, initially
preset with the current time and date. Using the
▲
key (14) and the
▼
key
(15), you can switch between date and time displays.
To change the settings for day / month / year or hours / minutes / seconds,
press the program key below and change the value using the
▲
key (14) or
the
▼
key (15).
or
P
1
P
1
P
2
P
2
P
3
or
or
day
hours
month
minutes
year
or
or
or
+
,
,
,
,
,
,
+
+
+
+
+
Cycle
30. 01. 07
Timer Date
Cycle
12h 20min
Timer Time
Figure 10: Setting the timer, with corresponding display
Example: Setting the hours
♦
Use the
▲
key (14) or the
▼
key (15) to change to the Timer display.
♦
Press and hold the P
1
program key (18).
♦
Press the
▲
key (14) or the
▼
key (15) until the desired hour is set.
♦
Release both keys (18 and 14 or 15).
Proceed in the same way to set the minutes and, if required, the date.
When you are done setting the timer, you must confirm your setting by
pressing the Timer key (23). This also activates the timer:
♦
Press the Timer key (23).
The cycle display (16) now shows the timer setting (Figure 12) with the
current time and date on the first line (U) and the timer setting on the second
line (T).
The timer is active when the indicator in the Timer key (24) is lit.
Sterilization does not start immediately when the START/STOP key (20) is
pressed, but on the day and at the time selected.
Zyklus
U 12 12.01.07
22
T 6 13.01.07
00
Figure 11: Timer display