5 Preparing for sterilization
03/2007
25
DA0621_a/
5.2.3 Loading the sterilizing chamber
♦
Load the items to be sterilized into the sterilizing chamber.
♦
Close the door (2) until it touches the sterilizing chamber (8).
Later, the door closes automatically as soon as a program is started.
Ö
When sterilizing liquids:
For sterilizing liquids, a reference temperature must be measured. If the sterilizer is
not equipped with a flexible temperature sensor, the temperature will be measured
automatically.
With
media temperature control expansion option
: The media temperature as
the reference temperature is measured using the flexible temperature sensor.
The following points should be observed in connection with the reference flask:
•
It must be at least as large as the vessels for the sterilizing articles.
•
It must be filled at least as far as the vessels for the sterilizing articles.
•
It must be placed in the back of the tray area.
The reference flask is filled with the liquid destined for sterilization. If this is not
possible, you may fill the reference flask with deionized water.
♦
Insert the flexible temperature sensor into the reference flask. The tip of
the temperature sensor must be in the bottom third of the reference flask
(minimum insertion depth: 20 mm).
♦
Place the reference flask in the back of the tray area.