7
Troubleshooting
Mechanical emergency lid release
7-2
SL 16 / 16R
Thermo Scientific
Mechanical emergency lid release
During a power failure, you will not be able to open the centrifuge lid with the regular electric lid release.
A mechanical override is provided to allow sample recovery in the case of an emergency. However, this
should be used only in emergencies and after the rotor has come to a complete stop.
Always wait a few minutes until the rotor has come to a stop without braking. The brake does not
work when there is no current. The braking process lasts much longer than usual.
Proceed as follows:
1. Make sure the rotor has stopped (view port in the lid).
2. Pull out the power supply plug.
3. On the left side of the housing is one white plastic plugs which you can pry out of the side plate
with a screwdriver or a knife.
Pull the release cord attached to it to trigger the mechanical lid release. The lid will open and the
samples can be removed.
Figure 7-6.
Emergency lid release
4. Push the cord back into the centrifuge and mount the plugs.
Reconnect the centrifuge once the power has been restored. Switch on the centrifuge. Press the
OPEN key
to have the lid locks operative again.
WARNING
The rotor can still be spinning at high speed. If touched, it can cause serious
injuries.
WARNING
Never use your hand or other tools to brake the rotor.
Emergency lid release
Содержание Thermo Scientific SL16
Страница 3: ......
Страница 21: ...Thermo Scientific SL 16 16R 3 1 3 Control panel Control panel on page 3 2 on page 3 3 ...
Страница 24: ...3 Control panel Control panel 3 4 SL 16 16R Thermo Scientific ...
Страница 34: ...4 Operation Switching the centrifuge off 4 10 SL 16 16R Thermo Scientific ...
Страница 42: ...6 Maintenance and care Service of Thermo Fisher Scientific 6 6 SL 16 16R Thermo Scientific ...
Страница 56: ...B Warranty B 2 SL 16 16R Thermo Scientific ...
Страница 60: ...Index I 2 SL 16 16R Thermo Scientific Temperature adaptation 4 8 Transporting the centrifuge 2 2 ...
Страница 61: ......