4
Operation
Rotor installation
Thermo Scientific
SL 16 / 16R
4-3
The rotors are equipped with an AutoLock
TM
locking system.
This system is used to automatically lock the rotor to the centrifuge spindle. The rotor does not have to
be bolted onto the centrifuge spindle.
Proceed as follows:
1. Open the lid of the centrifuge and if necessary remove any dust, foreign objects or residue from the
chamber.
AutoLock
TM
and o-ring must be clean and undamaged.
Figure 4-4.
AutoLock
TM
2. Hold the rotor over the shaft and let it slide slowly down the centrifuge spindle.
The rotor clicks automatically into place.
3. Check if the rotor is properly installed by lifting it slightly on the handle. If the rotor can be pulled
up, then it must be reclamped to the centrifuge spindle.
4. Place the rotor lid onto the rotor.
5. Close the centrifuge lid.
CAUTION
Do not force the rotor onto the centrifuge spindle.
If the rotor is very light, then it may be necessary to press it onto the centrifuge spindle
with a small amount of pressure.
WARNING
If the rotor cannot be properly locked in place after several attempts, then the
AutoLock
TM
is defective and you are not permitted to operate the rotor.
Check for any damage to the rotor: Damaged rotors must not be used.
Keep the hub area clear of objects.
CAUTION
Check that the rotor is properly locked on the centrifuge spindle before each
use by pulling it a its handle.
Be sure to check all sealings before starting any aeorsol-tight applications.
See the information in the rotor instruction manual.
AutoLock
TM
O-ring
Содержание Thermo Scientific SL16
Страница 3: ......
Страница 21: ...Thermo Scientific SL 16 16R 3 1 3 Control panel Control panel on page 3 2 on page 3 3 ...
Страница 24: ...3 Control panel Control panel 3 4 SL 16 16R Thermo Scientific ...
Страница 34: ...4 Operation Switching the centrifuge off 4 10 SL 16 16R Thermo Scientific ...
Страница 42: ...6 Maintenance and care Service of Thermo Fisher Scientific 6 6 SL 16 16R Thermo Scientific ...
Страница 56: ...B Warranty B 2 SL 16 16R Thermo Scientific ...
Страница 60: ...Index I 2 SL 16 16R Thermo Scientific Temperature adaptation 4 8 Transporting the centrifuge 2 2 ...
Страница 61: ......