1
Introduction and Description
Technical Data
Thermo Scientific
SL 16 / 16R
1-3
Technical Data
The technical data of the Thermo Scientific SL16 / 16R is listed in the following table.
Table 1-1.
Technical Data Thermo Scientific SL16 / 16R
Feature
Value
Environmental conditions
- use in interior spaces
- altitudes of up to 2,000 m above sea level
- max. relative humidity 80% up to 31 °C; decreasing linearly up to 50%
relative humidity at 40 °C.
permissible ambient temperature
+2° C to +35° C
+2° C to +35° C
Overvoltage category
II
II
Pollution degree
2
2
Heat dissipation
refrigerated
ventilated
230V
120V
230V
120V
4778 BTU/h
4096 BTU/h
3447 BTU/h
3901 BTU/h
IP
20
20
running time
unlimited
unlimited
maximum speed n
max
15200 rpm
15200 rpm
minimum speed n
max
300 rpm
300 rpm
maximum RCF value at n
max
25701xg 25701xg
maximum kinetic energy
<62,5 kJ
<51,7 kJ
noise level at maximum speed
<63 dB(A)
<63 dB(A)
temperature setting range
-10 °C to +40 °C
Dimensions
refrigerated
ventilated
Height
Height lid open
Width
Depth
Table top height
360 mm
870 mm
623 mm
605 mm
310 mm
360 mm
870 mm
440 mm
605 mm
310 mm
weight without rotor
91,5 kg
57,5 kg
Содержание Thermo Scientific SL16
Страница 3: ......
Страница 21: ...Thermo Scientific SL 16 16R 3 1 3 Control panel Control panel on page 3 2 on page 3 3 ...
Страница 24: ...3 Control panel Control panel 3 4 SL 16 16R Thermo Scientific ...
Страница 34: ...4 Operation Switching the centrifuge off 4 10 SL 16 16R Thermo Scientific ...
Страница 42: ...6 Maintenance and care Service of Thermo Fisher Scientific 6 6 SL 16 16R Thermo Scientific ...
Страница 56: ...B Warranty B 2 SL 16 16R Thermo Scientific ...
Страница 60: ...Index I 2 SL 16 16R Thermo Scientific Temperature adaptation 4 8 Transporting the centrifuge 2 2 ...
Страница 61: ......