Before use
Aligning the centrifuge
Thermo Scientific
SL 16 / 16R
2-3
•
The supporting structure must be stable and free of resonance, for example a level laboratory
bench.
•
The supporting structure must be suitable for horizontal setup of the centrifuge.
•
The centrifuge should not be exposed to heat and strong sunlight.
•
The set-up location must be well-ventilated at all times.
Aligning the centrifuge
The alignment of the centrifuge must be checked every time after moving it to a different location.
The supporting structure must be suitable for horizontal setup of the centrifuge.
Mains connection
Figure 2-2.
Mains connections
1. Turn off the power supply switch on the back (press "0").
2. Plug the centrifuge into grounded electrical sockets only.
3. Check whether the cable complies with the safety standards of your country.
WARNING
UV rays reduce the stability of plastics.
Do not subject the centrifuge, rotors and plastic accessories to direct sunlight.
VORSICHT
If the centrifuge isn’t level, imbalances can occur and the centrifuge can be
damaged.
Do not place anything under the feet to level the centrifuge.
Main Power Switch
Mains plug
Содержание Thermo Scientific SL16
Страница 3: ......
Страница 21: ...Thermo Scientific SL 16 16R 3 1 3 Control panel Control panel on page 3 2 on page 3 3 ...
Страница 24: ...3 Control panel Control panel 3 4 SL 16 16R Thermo Scientific ...
Страница 34: ...4 Operation Switching the centrifuge off 4 10 SL 16 16R Thermo Scientific ...
Страница 42: ...6 Maintenance and care Service of Thermo Fisher Scientific 6 6 SL 16 16R Thermo Scientific ...
Страница 56: ...B Warranty B 2 SL 16 16R Thermo Scientific ...
Страница 60: ...Index I 2 SL 16 16R Thermo Scientific Temperature adaptation 4 8 Transporting the centrifuge 2 2 ...
Страница 61: ......