7
Troubleshooting
Troubleshooting by user
7-4
SL 16 / 16R
Thermo Scientific
Imbalanced load
The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
Imbalance detected.
Check the load placed in the rotor.
Check that the rotor cross bolts are well greased.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
Check Set Speed
The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
The set point speed is higher than the maximum rotor speed.
Correct the value.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
E-01 - E-12
The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
Error during the self-test of the centrifuge program and the
electronics.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
E-13
The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
The check sum in the data memory is incorrect.
The software corrects errors automatically.
Check the values of the set point settings, etc.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
E-15 - E-16
Temperature sensor broken / controller
defective
Malfunction in the temperature detection.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
E-17
Speed for rotor detection exceeded
Restart the centrifuge.
If an error message appears again, inform Customer Service.
E-21 - E-22
The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
The rotor could not be identified.
Check to see if the rotor is properly installed.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
E-23
The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
The speed control measurement returned a different result.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
E-25-E-27
The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
Time has expired for the lid lock drive while opening the lid.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
E-28
The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
Error during the self-test of the centrifuge program and the
electronics.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
Failure message
Problem with centrifuge
Possible causes and cures
Содержание Thermo Scientific SL16
Страница 3: ......
Страница 21: ...Thermo Scientific SL 16 16R 3 1 3 Control panel Control panel on page 3 2 on page 3 3 ...
Страница 24: ...3 Control panel Control panel 3 4 SL 16 16R Thermo Scientific ...
Страница 34: ...4 Operation Switching the centrifuge off 4 10 SL 16 16R Thermo Scientific ...
Страница 42: ...6 Maintenance and care Service of Thermo Fisher Scientific 6 6 SL 16 16R Thermo Scientific ...
Страница 56: ...B Warranty B 2 SL 16 16R Thermo Scientific ...
Страница 60: ...Index I 2 SL 16 16R Thermo Scientific Temperature adaptation 4 8 Transporting the centrifuge 2 2 ...
Страница 61: ......