Maintenance
2. Remove the source.
a. Carefully tip the instrument back until it rests on the electronics enclosure.
b. Use a #1 Phillips head screwdriver to fully loosen the three captive screws that hold the source in
place.
c. Swing the bale out. Steady the instrument with one hand and pull firmly on the bale to pull the
source straight out of the instrument.
54
Nicolet Summit LITE FTIR Spectrometer User Guide
Thermo Scientific