Options and Accessories
WARNING
Avoid explosion hazard.
Before opening the desiccant compartment, turn off the instrument power,
unplug the power cord and remove all accessories and samples from the
system. Entry of flammable liquids or gases into the desiccant
compartment is an explosion hazard. If such entry occurs, contact us
immediately and do not apply power to the instrument until the condition
has been corrected.
b. Loosen the two captive Phillips head screws from the standard desiccant cartridge, and lift the
cartridge straight up and out of the instrument.
Note If the standard desiccant cartridge is stored in a dry box or sealed bag (you can use the
bag the purge cartridge came in), the cartridge can be reused.
c. Remove and discard the large O-ring.
d. Open the purge kit and the sealed desiccant packet.
36
Nicolet Summit LITE FTIR Spectrometer User Guide
Thermo Scientific