To replace the desiccant
1. Turn off the spectrometer.
a. To turn off the spectrometer, press the power button.
b. Unplug the power cord.
c. Turn off the purge, if applicable, and disconnect the purge line to the instrument and any
accessory.
Note You can use either desiccant cartridge (with or without the purge connector) to maintain
the required humidity. The illustrations below show the nonpurged version of the desiccant
cartridge. The instructions for replacing the desiccant canisters are the same for both
cartridges.
2. Remove the desiccant cartridge.
a. Remove any installed accessory and all samples from the instrument.
b. Use the #2 Phillips head screwdriver to loosen the two captive screws from the desiccant
cartridge, and lift the cartridge straight up and out of the instrument.
Thermo Scientific
Nicolet Summit LITE FTIR Spectrometer User Guide
49
Maintenance