Unpack and Get Started with the
Spectrometer
Unpacking and setting up your Thermo Scientific Nicolet Summit Spectrometer requires you to place the
instrument in your workspace, to connect and power on the spectrometer, and to run a short
performance test.
Before You Begin
NOTICE
Wait 24 hours before taking the spectrometer out of its plastic shipping bag to prevent condensation
from forming that can damage internal optics. The warranty will not cover damage due to opening the
sealed plastic bag before the instrument has come to room temperature.
Reminder: Review the spectrometer's Site and Safety Information to ensure that the workspace is safe
and suitable for the instrument.
Unpacking and Setup
The steps for unpacking and setting up your spectrometer vary somewhat depending on the options you
purchased.
1. Unpack the shipping box
Remove the spectrometer from the shipping box and bag and remove additional packing materials.
NOTICE
Use care when lifting the spectrometer.
l
Do not touch the sample compartment windows while unpacking the instrument. Touching the
windows could cause permanent damage.
Thermo Scientific
Nicolet Summit LITE FTIR Spectrometer User Guide
15
Get Started with Your Nicolet Summit Spectrometer