CAUTION
Avoid burn hazard.
Internal components, especially the infrared source, can become extremely hot
during normal operation. Turn off the instrument and wait at least 10 minutes
before replacing any components.
WARNING
Avoid personal injury.
l
Never stare into the laser beam or at its reflection. Never tamper with the laser.
You could be exposed to laser light or high voltage.
l
If you adjust the laser or perform procedures that are not described in the user
guides and manuals, you could be exposed to hazardous radiation.
REPLACING THE DESICCANT
When you open the desiccant compartment, you must prevent flammable liquids or gases from entering
the compartment. For instructions on changing the desiccant, see
.
DANGER
Avoid explosion hazard.
Before you open the desiccant compartment, turn off the instrument power, unplug
the power cord and remove all accessories and samples from the system. Entry of
flammable liquids or gases into the desiccant compartment could cause an
explosion. If such entry occurs, contact us immediately and do not apply power to
the instrument until the condition has been corrected.
REPLACING THE IR SOURCE
The IR source becomes extremely hot during use. If you need to replace the IR source, avoid burn and
explosion hazards.
CAUTION
Avoid burn hazard.
The source becomes extremely hot during normal use. Always allow the source to
cool for at least 10 minutes after the spectrometer is turned off before you work
with the source.
Thermo Scientific
Nicolet Summit LITE FTIR Spectrometer User Guide
13