Maintenance
ii. Select Start to begin alignment. The screen says "Align complete" when alignment has fin-
ished successfully.
b. Calibrate the laser.
i. Go to Acquire Data > Diagnostics > Laser Calibration.
ii. Select Start to begin calibration.
Laser calibration may take several minutes to complete. When calibration is finished, the
display shows "Measurement complete" and indicates any change in laser frequency.
The instrument takes up to 12 hours to come to thermal equilibrium. For best results, re-run
alignment and calibration after the first 12 hours of operation.
c. Run thet Nicolet FTIR – Factory Qualification workflow. Performance verification and
qualification workflows use a polystyrene reference material inside the spectrometer to run a
series of standard tests to verify your instrument operation and to ensure the accuracy of your
data.
i. Return to the Dashboard.
ii. Scroll to the Workflows pane and select Nicolet FTIR - Factory Qualification.
iii. Right click the workflow and select Run to start the workflow.
6. Replace any accessories that you removed earlier.
56
Nicolet Summit LITE FTIR Spectrometer User Guide
Thermo Scientific