Options and Accessories
Hardware
Pressure
(psig)
Flow Rate (scfh)
Summit Spectrometer
5
1
Everest Accessory
5
4-10
iD1 Transmission
Accessory
5
4-10
iD3 ATR Accessory
5
4-10
iD5 ATR Accessory
5
4-10
iD7 ATR Accessory
5
4-10
iD Foundation
5
4-10
b. Plug in the power cord to the instrument and turn on the power.
c. Wait 30 - 60 minutes for the instrument to fully purge.
Set Purge Gas Controls
Correctly setting your purge gas controls will protect your spectrometer from humidity without
introducing vibration.
For best results, the purge gas should be dried to a dew point of -70 °C (-94 °F) or below.
❖
To set the purge gas controls
1. Open the main valve to start the flow of purge gas through the regulator.
2. Adjust the pressure regulator for the spectrometer until the gauge indicates that the pressure is 5
psig (34 kPa).
38
Nicolet Summit LITE FTIR Spectrometer User Guide
Thermo Scientific