Operation
Calibrating the Micro-Tech
3-10
Micro-Tech 9101/9201 Reference Manual, Rev J
Thermo Fisher Scientific
4.
Make sure the belt is running empty at maximum speed, then press
the Start button. The count-down screen appears. (The data in your
screen will, of course, be different.)
NOTE.
The number of seconds shown in “Time remaining” is
calculated based on the current speed-sensor pulse frequency, and
estimates the time remaining for a complete test.
5.
The calibration time (in seconds) that you established during the
Micro-Tech cold-start procedure, will start counting down. When the
counter reaches zero, the calibration is complete and the change-span
screen appears.
BATCH
AUTO SPANNING
Time remaining 375
Rate 150 Tph
Tot 0.000 Tons
READY
ALARM
CALIB
ABORT
BATCH
AUTOSPAN COMPLETE
Change span?
Error 0.01%
READY
ALARM
CALIB
NO
YES
ADV