Operation
Calibrating the Micro-Tech
Thermo Fisher Scientific
Micro-Tech 9101/9201 Reference Manual, Rev J
3-9
To perform an R-Cal span calibration for your scale, do the following.
1.
Make sure the Run screen (see
) is currently being
displayed.
2.
Press the Menu button and the “Main Menu 1” screen appears.
3.
Press the Span Calibration button and the following screen appears.
If the R-Cal auto-span option (shown above) is
not
displayed, go to
page B-5 to change the span calibration method. (For additional
information about the Calibration menu, go to page B-22.)
Doing an R-Cal
Span Calibration
BATCH
MAIN MENU 1
Press MENU for more
READY
ALARM
CALIB
SPAN CAL
ZERO CAL
MATL CAL
BATCH
AUTO SPAN R Cal
Run the belt empty, then
press START
READY
ALARM
CALIB
EXIT
START
MANUAL