B-28
Micro-Tech 9101/9201 Reference Manual, Rev J
Thermo Fisher Scientific
No. Parameter
Selection
Description
22
Zero Ready
no, yes
The Zero Ready feature provides visible
indication of when the scale is not loaded,
has a valid zero calibration, and is ready to
run material. This feature is a requirement
for all NTEP scales. The indication is shown
on the display when the belt loading remains
under 0.12% of full scale belt loading for a
period of time equivalent to one belt
revolution. The indicator is the green ZERO
LED above the display. The function can be
activated only after a complete scale
calibration. A Zero Ready digital output is
also available, it is active when the green
ZERO LED is displayed.
When W&M Mode is set to NTEP, the Zero
Ready scroll defaults to “Yes”. When W&M
Mode is set to OIML or NONE, the Zero
Ready scroll defaults to “No”.
Whenever the Zero Ready scroll setting is
changed an Audit Trail event is generated.
(Password: Service.)
Содержание Micro-Tech 9101
Страница 1: ...Micro Tech 9101 9201 Integrator Reference Manual REC 4281 Rev J Part Number 127336 English...
Страница 82: ......
Страница 118: ......
Страница 124: ...A 2 Micro Tech 9101 9201 Reference Manual Rev J Thermo Fisher Scientific...
Страница 172: ......
Страница 234: ......
Страница 315: ......
Страница 316: ......
Страница 317: ......
Страница 318: ......
Страница 319: ......
Страница 320: ......
Страница 321: ......
Страница 322: ......
Страница 323: ......
Страница 324: ......
Страница 325: ......
Страница 326: ......
Страница 327: ......
Страница 328: ......
Страница 329: ......