Thermo Fisher Scientific
Micro-Tech 9101/9201 Reference Manual, Rev J
B-61
Here is the Linearization menu table.
No. Parameter
Selection
Description
1
Linearization
no, yes
Note: No linearization is done until you perform
an Auto Multipoint material calibration or
manually enter the linearization factors.
(Password: Service.)
LINEARIZ. #1
Load ______
Fact. ________
0.00 Lb/Ft
1.000000 (0-1.5)
This scroll is for flow rate under 20%.
Enter a Factor and a Load.
If you enter 0.000000 for the factor, the rate is
not corrected in that portion of the range. A
factor lower than 1.000 reduces the span, while
a factor larger than 1.000 increases the span.
The ACQ key automatically acquires the Load
at the present flow rate. (Password: Service.)
LINEARIZ. #2
Load ______
Fact. ________
0.00 Lb/Ft
1.000000 (0-1.5)
This scroll is for flow rate under 40%.
"
LINEARIZ. #3
Load ______
Fact. ________
0.00 Lb/Ft
1.000000 (0-1.5)
This scroll is for flow rate under 60%.
"
LINEARIZ. #4
Load ______
Fact. ________
0.00 Lb/Ft
1.000000 (0-1.5)
This scroll is for flow rate under 80%.
"
LINEARIZ. #5
Load ______
Fact. ________
0.00 Lb/Ft
1.000000 (0-1.5)
This scroll is for flow rate under 100%.
"
Содержание Micro-Tech 9101
Страница 1: ...Micro Tech 9101 9201 Integrator Reference Manual REC 4281 Rev J Part Number 127336 English...
Страница 82: ......
Страница 118: ......
Страница 124: ...A 2 Micro Tech 9101 9201 Reference Manual Rev J Thermo Fisher Scientific...
Страница 172: ......
Страница 234: ......
Страница 315: ......
Страница 316: ......
Страница 317: ......
Страница 318: ......
Страница 319: ......
Страница 320: ......
Страница 321: ......
Страница 322: ......
Страница 323: ......
Страница 324: ......
Страница 325: ......
Страница 326: ......
Страница 327: ......
Страница 328: ......
Страница 329: ......