A-11
7.
Press the Count button each time a reference mark passes.
8.
Continue to press the Count button when a mark passes, until the
following two conditions are met.
The belt has made at least
three
complete revolutions.
The total running time of the test exceeds
six
minutes.
9.
After the last revolution has been counted, press the Done button.
BATCH
Press COUNT each
time a mark passes.
1 sec 0 rev
READY
ALARM
CALIB
ABORT
COUNT
DONE
BATCH
Press COUNT each
time a mark passes.
120 sec 1 rev
READY
ALARM
CALIB
ABORT
COUNT
DONE
BATCH
TEST DURATION
Length = 600.0 ft
Time = 360 sec
READY
ALARM
CALIB
EXIT