Standardization & Calibration
Perform Calibration
3.
Fill the tank with process material to a particular level you are
interested in. Measure the level accurately and keep it steady as
possible.
4.
In the NI-FBUS Configurator, go to
AI_TRANSDUCER > Root
Menu > LevelPro setup > 2 Gauge Calibration > GAUGE
CALIBRATION.
5.
The first calibration screen appears. Click
Next
when you are ready to
continue.
Figure 5–6.
Initial gauge calibration screen
6.
At the prompt, specify the number of time constant periods used for
the standardization. If the time constant is 128 seconds and you enter
2
here, the standardization cycle will last 256 seconds (2 x 128 seconds).
Click
Next
.
Thermo Fisher Scientific
LevelPRO with
F
OUNDATION
™ fieldbus
Application Guide
5-5
Содержание LevelPRO Series
Страница 2: ......
Страница 3: ...LevelPRO With FOUNDATION TM Fieldbus Application Guide P N 717915 Revision A ...
Страница 4: ......
Страница 6: ...This page intentionally left blank ...
Страница 8: ...This page intentionally left blank ...
Страница 54: ...This page intentionally left blank ...
Страница 64: ...This page intentionally left blank ...
Страница 70: ...This page intentionally left blank ...
Страница 94: ...This page intentionally left blank ...
Страница 98: ...This page intentionally left blank ...