LevelPRO Setup
Time Constants
Figure 4–6.
Calibration point #1
Time Constants
Note
For detailed discussion on time constants, reference the LevelPRO
user guide.
▲
To set the primary and alternate time constants using the NI-FBUS
Configurator:
1.
Go to
AI_TRANSDUCER > Root Menu > LevelPro setup > 1 Main
Measurement > 4 Time Constant > Time constant setup
.
2.
The first time constant setup screen will appear. Click
Next
when ready
to continue.
Thermo Fisher Scientific
LevelPRO with
F
OUNDATION
™ fieldbus
Application Guide
4-5