Standardization & Calibration
Perform Standardization
4.
The first standardization screen appears. Click
Next
when you are
ready to continue.
Figure 5–1.
Initial standardization screen
5.
At the prompt, specify the number of time constant periods used for
the standardization. If the time constant is 128 seconds and you enter
2
here, the standardization cycle will last 256 seconds (2 x 128 seconds).
Click
Next
.
Figure 5–2.
Setting the number of time constants used for standardization
5-2
LevelPRO with
F
OUNDATION
™ fieldbus
Application Guide
Thermo Fisher Scientific