Sarasota 2000 Ultrasonic Multipath
Flowmeter
Appendix 1 SCREEN FLOW DIAGRAMS, Page A1-44
Thermo Fisher Scientific
APPENDIX 1 LCD SCREENS
SCREEN C22: ULTRASONIC VELOCITY PATHS (MENU)
(Screen accessed via C2:Channel Setup; ESC C220:Path Menu)
C22 Ultr asoni c Pat hs
Path 1 [ A1] I F Car d 1
Path 2 [ A2] I F Car d 1
Path 3 [ B1] I F Car d 1
Path 4 [ B2] I F Car d 1
S tatio n C lone Reb uild
ULTRASONIC PATH ALLOCATION
•
Screen available for each configured channel
•
See
Section
2.3.3
•
Paths shown have been selected via screen S230:Path Allocation
•
In the example above, the transducer for paths 1 and 2 are multi-dropped and
connected to rear terminal A. Similarly, paths 3 and 4 are multi-dropped and
connected to rear terminal B.
•
Highlighting the selected path and pressing ENTER accesses C220:Path Menu to
enable configuration of each path
•
Alternatively, a ‘cloning’ function is available via the Clone soft key as a short cut for
applications where path configuration is similar (eg. a vertical sided channel). The
configuration of the highlighted path is applied to all paths below in the list, however
path height will still need to be entered via screen C221:Path Configuration Edit.
•
On completion of configuration, press Rebuild soft key