Sarasota 2000 Ultrasonic Multipath
Flowmeter
Appendix 1 SCREEN FLOW DIAGRAMS, Page A1-43
Thermo Fisher Scientific
APPENDIX 1 LCD SCREENS
SCREEN C218: WAVEFORM DISPLAY (WAVEFORM)
(Screen accessed via C217:Waveform Capture Setup)
Tof:0.637mS
0.600mS Level W1 1.112mS
level
WAVEFORM DISPLAY
•
Screen available for each configured ultrasonic level input
•
Time of flight (tof) value typically taken as trigger value input for screen
C217:Waveform Capture Setup
•
Detection levels shown as horizontal dotted lines.
•
Initial number shown at bottom left of screen is trigger delay. Initial number shown at
bottom right of screen is duration (trigger delay + sample period)
•
On access to C218:Waveform Display - Level from C217:Waveform Capture Setup,
the initial waveform relates to sample period input in C217:Waveform Capture Setup
•
UP and DOWN arrow keys can be used to zoom in and out and LEFT and RIGHT
arrow keys can be used to shift the display window