Sarasota 2000 Ultrasonic Multipath
Flowmeter
Section 7, MAINTENANCE Page 7-2
Thermo Fisher Scientific
MAINTENANCE
7.3 Routine
checks
7.3.1 Remote
Remote checks of operation of all velocity paths and depths are easily carried out via GAFA.
It is recommended that a monthly routine should be set up to handle this.
If the flowmeter is fitted with the relay option, one of the relays may be programmed to initiate
an early warning call via a telemetry outstation.
7.3.2 On
site
It is recommended that an annual check be made on site. This should include:
•
A visual observation of the equipment in the channel and the flowmeter.
•
A functional check of the operation of the paths, levels and outputs.
•
A check on the operation of the internal battery.
•
A verification of calculated flow by a comparison method, see Section 6: Calibration
Any corrective work should be taken at the time if possible or reported for subsequent action.