![Thermo Scientific HB-S2000 Скачать руководство пользователя страница 59](http://html.mh-extra.com/html/thermo-scientific/hb-s2000/hb-s2000_user-manual_1105188059.webp)
Sarasota 2000 Ultrasonic Multipath
Flowmeter
Appendix 1 SCREEN FLOW DIAGRAMS, Page A1-11
Thermo Fisher Scientific
APPENDIX 1 LCD SCREENS
SCREEN C10: INSTANTANEOUS FLOW & LEVEL (RESULTS)
(Screen accessed via C1:Channel Diagnostics)
C10 Instantaneous Flow & Level
Flow : 1.60 litres/sec
DMD : 0.631 litres/sec
Arbitrated Level: 9.72 metre
i/p Type Status
Level
1L USW1 ok
10.45
metre
2L BCD1 ok
9.89
metre
3L ENC1 ok
9.55
metre
4L
Aux
Aux
Aux
Aux
Channel1
INSTANTANEOUS FLOW AND LEVEL
•
Screen available for each configured channel
•
For ultrasonic depths (USW): status
can be ‘OK’; ‘no signal’, ‘no transducer’, ‘too low’
and ‘no vos’
•
For 4-20mA analogue (AIN): status can be ‘OK’ or ‘no signal’
•
For binary coded decimal (BCD): status can be ‘OK’ or ‘no signal’
•
For encoder (ENC) auxiliary depths: status can be ‘OK’ or ‘no signal’
•
For arbitration rules see section 1.3:
•
Auxiliary levels (1-4) are displayed for all channels. Auxiliary levels are not used to
calculate the channel level and are not affected by arbitration rules.
•
These results are instantaneous – see Section 1.3
SCREEN C11: INSTANTANEOUS PATH VELOCITIES (RESULTS)
(Screen accessed via C1:Channel Diagnostics)
C11 Instantaneous Path Velocities
Velocity Units: metres/sec
mS uS
Path Status
Vos Vel tof dif
1 100%
1507
0.027
0.631
0.020
2 100%
1496
0.002
0.615
0.001
Channel1
INSTANTANEOUS PATH VELOCITIES
•
These results are instantaneous - see Section 1.3 for definition
•
See Section 1.3 for details about Status percentage reading