4
– Applications
Thermo Scientific
Product Manual for CarboPac PA200 Columns
Page 21 of 37
065734-01
For Research Use Only. Not for use in diagnostic procedures.
0-50 min: 15-64 mM KMSA in 136 mM KOH
50-60 min: 80 mM KMSA in 90 mM KOH
60-65 min: 100 mM KMSA in 100 mM KOH
65-80 min: 15 mM KMSA in 136 mM KOH
Flow Rate:
CarboPac PA200, 3-mm: 0.5 mL/min
CarboPac PA200, 1-mm: 0.063 mL/min
CarboPac PA200, 0.4-mm: 0.010 mL/min
Detection:
CarboPac PA200, 3-mm: PAD, Au on PTFE, 2 mil gasket, Ag/AgCl ref.
CarboPac PA200, 1-mm: PAD, Au on PTFE, 1 mil gasket, Ag/AgCl ref.
CarboPac PA200, 0.4-mm: PAD, Au on PTFE, 1 mil gasket, Ag/AgCl ref.
Samples:
50 µmol/L fetuin oligosaccharide alditol standard
Figure 3
Fetuin Oligosaccharide Profiles: CarboPac PA200, 3-mm vs. 1-mm vs. 0.4mm